Both PIM and S-parameters, such as return loss or insertion loss, are essential measurements for passive components. Both must be tested in order to fully qualify a passive device. Conventional test systems require a network analyzer and separate PIM analyzers to test a device completely. If different test stations are used, the time required to connect and disconnect components can significantly exceed actual testing time.
AWT’s new integrated solution PEM42A allows you to measure PIM and S-parameters with a single DUT connection. It utilizes the Agilent E5072A ENA series network analyzer to perform S-parameter measurements together with AWT’s PEM42A multi band PIM analyzers. Whether for R&D, production or QA , this system is well suited for testing wideband components.
Conventional PIM analyzers are designed for a specific frequency band. Testing wide band devices requires several PIM analyzers. PEM42A multi band systems combine up to four frequency bands to one unified system. All instruments are controlled by a PEM42ACTRL Control & Switch Master Unit for synchronized measurements. PEM42A systems are a cost-effective solution, providing very fast and accurate measurements. Repeated connect / disconnect cycles for measuring PIM of DUTs in different frequency bands is a thing of the past. Both PIM and S-parameters are now measured without changing the physical connection. This results in a major reduction of overall test time and increases test throughput significantly.
Model Number |
Type |
Tx Frequency (MHz) |
Rx Frequency (MHz) |
PEM42A-CTRL | Master Control & Switch Unit (controls up to 4 PEM42A RF Modules) |
All | All |
RF Units |
|||
PEM42A-385 | TETRA 385 | 385 ~ 400 | 410 ~ 425 |
PEM42A-400 | TETRA 400 | 390 ~ 400 | 380 ~ 385 |
PEM42A-400E | E-TETRA | 420 ~ 430 | 410 ~ 412 |
PEM42A-700L | LTE 700 (Lower Band) | 728 ~ 759 | 698 ~ 716 |
PEM42A-700U | LTE 700 (Upper Band) | 730 ~ 759 | 776 ~ 788 |
PEM42A-800 | LTE 800 | 791 ~ 821 | 832 ~ 862 |
PEM42A-850 | AMPS / CDMA | 869 ~ 896 | 824 ~ 851 |
PEM42A-900E | E-GSM 900 & GSM900 | 925 ~ 960 | 880 ~ 915 |
PEM42A-1800 | DCS 1800 /GSM 1800 | 1805 ~ 1880 | 1710 ~ 1785 |
PEM42A-1900 | PCS 1900 | 1930 ~ 1990 | 1850 ~ 1910 |
PEM2000TD | TD-SCDMA(2000) | 2010 ~ 2025 | 1900 ~ 1920 |
PEM2100AWS | AWS | 2010 ~ 2155 | 1710 ~ 1755 |
PEM2100UMTS | UMTS/W-CDMA | 2110 ~ 2170 | 1920 ~ 1980 |
PEM2160JP | W-CDMA-JP | 2150 ~ 2170 | 2110 ~ 2140 |
PEMWBKR | WiBro-KR | 2110 ~ 2170 2300 ~ 2390 |
1910 ~ 1990 |
PEM2600LTE | IMT-E(2600), LTE2600 | 2620 ~ 2690 | 2500 ~ 2570 |
Agilent |
|||
PEM42A-E5072A-245 | Agilent Network Analyzer, 2-port, 30k to 4.5 GHz, configurable test set, opt 008: frequency offset mode, opt 1E5: high stability timebase, opt 1CP: rack mount kit. Prices and delivery might vary and dependent on Agilent policies. Sold only with PEM42A PIM analyzers. | ||
PEM42A-E5072A-285 | Agilent Network Analyzer, 2-port, 30k to 8.5 GHz, configurable test set, opt 008: frequency offset mode, opt 1E5: high stability timebase, opt 1CP: rack mount kit. Prices and delivery might vary and dependent on Agilent policies. Sold only with PEM42A PIM analyzers. | ||
PEM42A-U2001A | Agilent USB power sensor, 10 MHz – 6 GHz. Required for user calibration of E5072A network analyzers. Prices and delivery might vary and dependent on Agilent policies. Sold only with PEM42A PIM analyzers. | ||
PEM42A-85032A | Agilent mechanical calibration kit for Network Analyzer E5072A. DIN 7/16, 50 Ohms, DC-7.5 GHz. Required for user calibration of E5072A network analyzers. Prices and delivery might vary and dependent on Agilent policies. Sold only with PEM42A PIM analyzers. | ||
PEM42A-85032F-100 | Agilent Network Analyzer, 2-port, 30k to 4.5 GHz, configurable test set, opt 008: frequency offset mode, opt 1E5: high stability timebase, Agilent mechanical calibration kit for Network Analyzer E5072A. Type-N with adaptor (f)-(f), 50 Ohms, DC-9GHz. Required for user calibration of E5072A network analyzers. Prices and delivery might vary and dependent on Agilent policies. Sold only with PEM42A PIM analyzers. | ||
Accessories |
|||
PLOAD100 | Low PIM Termination 100W, PIM < -165 dBc, 380 – 2800 MHz, Connector DIN(f) | ||
ADA-DMDF | Low PIM adapter, PIM < -165dBc @2x43dBm, DIN(m)-DIN(f), Connector Saver | ||
LIC308-DMDM-2M | Low PIM cable, Length: 2 meter (6 ft), PIM < -165dBc, DIN(m)-DIN(m) | ||
LIC308-DMDM-1M | Low PIM cable, Length: 1 meter (3 ft), PIM < -165dBc, DIN(m)-DIN(m) | ||
PACC2 | Accessory Kit: low PIM cable 3m DIN(m)-DIN(m), low PIM Load 50W / 10W, Frequency 698 to 2800 MHz, dual port DIN(m)-DIN(f), PIM <-165dBc dBm, torque wrench DIN7/16, (4) low PIM adaptors, cleaning tabs, hard carrying case IP66 | ||
PACC2L | Accessory Kit: low PIM cable 3m DIN(m)-DIN(m), low PIM Load 100W, Frequency 380 to 2800 MHz, DIN(f), PIM <-165dBc dBm, torque wrench DIN7/16, (4) low PIM adaptors, cleaning tabs, hard carrying case IP66 | ||
PIMGEN | PIM generator, preset PIM value 80 dBm (+/-10) for quick system tests. |
Transmitter |
|
Frequency Increments | 100 kHz |
Power Accuracy | +/- 0.65 dB |
Frequency Accuracy | 2 ppm |
Power (per tone) | 20 – 44 dBm adjustable |
– with PIM42AOPT5 | 20 – 46 dBm |
Reverse Pwr. Protection | +43 dBm for max 5 sec |
Adjustable Resolution | 0.25 dB |
Receiver |
|
Reverse IM @ 2x43dBmForward IM @ 2x43dBm | -122 dBm (Typ. -165 dBc @ 43 dBm x 2tones) |
Forward IM @ 2x43dBm | -120 dBm (Typ. – 163 dBc @43dBm x 2tones) |
Noise Floor | <-135 dBm typ |
DynamicRange(typical) | 125dB |
Reverse Power Protection | +43 dBm for 5 sec |
Max Input Power | -10 dBm |
Measurement Accuracy | +/- 1.0 dB Typ. |
Rack(With Auto-Caster) |
|
Dimensions | 600 x 1400 x 900 to 600 x 1533 x 1000 (mm) – 23.6 x 55.1 x 35.4 to 23.6 x 60.4 x 39.4 (in) |
Weight | 90 (kg) – 198.4 (lbs) |
Unit Dimensions |
|
PEM42A-CTRL | 483 x 222 x 664 (mm) – 19 x 8.7 x 26.1 (in) |
PEM42A | 483 x 133 x 657 (mm) – 19 x 5.2 x 25.9 (in) |
E5072A | 483 x 222 x 472 (mm) – 19 x 8.7 x 18.6 (in) |
Unit Weight |
|
PEM42A-CTRL | 29 (kg) – 63.9 (lbs) |
PEM42A | 25 to 28 (kg) – 55.1 to 61.7 (lbs) |
E5072A | 19.8 (kg) – 43.7 (lbs) |
Environmental/Electrical |
|
Operation Temperature | +5° C to +40° C |
HumidityStorageRelative | -10°C to +60°C / 20 to 80 % RH max (non-condensing) |
Power Consumption | PEM42A-CTRL & PEM42A 750 VA max.E5072A , 305 VA max |
Main Power | 100 to 240V 50/60 Hz |